Minimum detection limits and applications of proton and helium induced X-ray emission using transition-edge sensor array
2017
Abstract We have determined minimum detection limits, MDLs, for elements 14 ⩽ Z ⩽ 86 using a transition-edge sensor array, TES array, and as a comparison using an Amptek X-123SDD silicon drift detector, SDD. This was done using a 3 MeV proton beam and a 5.1 MeV helium beam. MDLs were determined for a thin film sample on top of C substrate, and for a bulk sample containing mostly Al. Due to the higher peak-to-background ratio, lower detection limits were obtainable using the TES array for most of the elements. However, for elements 30 ⩽ Z ⩽ 45 the performance of the TES array was not as good as the SDD performance. This is due to the limitations of the TES used at energies >10 keV. The great…